SPM-300 is a spirometer sensor which works with the Cardio7-S and SpiroCare. SPM-300 uses a disposable sensor, so it is very easy to maintain because the sensor does not require disinfecting between uses. These disposable sensors make an accurate measurement possible every time.
It is highly recommended to calibrate a spirometer with Bionet 3L calibration syringe periodically, in order to ensure accurate calibration of your spirometer.
The volume accuracy of the spirometer must be checked at least daily with a single discharge of a 3-L calibrated syringe.
At least quarterly, volume spirometers must have their calibration checked over their entire volume range using a calibrated syringe or an equivalent volume standard. The measured volume should be within 3.5% of the reading or 65 mL, whichever is greater. This limit includes the 0.5% accuracy limit for a 3-L syringe.
Spiro Handel SPM-300 Compatible with Cardio10
World patented spirometer handle.
By adding the SPM-300 to the Cardio7, you can test FVC, SVC, MVV and perform a pre-post bronchodilator comparison. Real-time volume-flow and time-flow monitoring is provided on the color LCD display. 120 tests are stored and a full size printout is provided for easy charting.
- Includes: main unit, nose clip, 1 adapter for reusable mouth pieces, 1 box (100 pcs) disposable mouth piece,
operation manual
- Pre- and post-bronchodilator comparison
- Dimension: 47 (W) x 200 (H) x 34 (D)mm, approx. 250g
Measuring Values:
- FVC: FVC, FEV 1.0, FEV 1/FVC, FEF 0.2 - 1.2L, FEF 25 - 75%, FEF 75 - 85%, PEF, FEF 25%, FEF 50%, FEF 75%, FIVC, FEV6, PEFT, FET 100%
- SVC: SVC, ERV, IRV, TV, EC
- MVV: MVV, FB, TV
Presentation:
- Flow volume loop
- MVV: MVV, FB, TV
- Measurement values table
Measuring Range
- Flow: 0 to ± 14 L/s
- Volume: 0 to ± 12 L
- Measuring Method: Differential pressure method
- Prediction Equation: Morris-Polgar, ECCS-Quanjer, Knudson-ITS, Pereira, Korea CJK
- Sample Rate: 200 samples/sec
- Flow Impedance: < 0.2 mbar s/l at 12 l/s
Environment:
- Ambient temperature: 15 to 40°C (59 to 104°F)
- Relative humidity: 10 to 90% (non-condensing)
- Atmospheric pressure: 700 to 1060 hPa
Standard Accessories:
- Dispoabla moutpiece - 100 ea
- Nose clip - 1 ea
- Adapter - 1 ea
- Handle dock - 1 ea
- Operation manual - 1 ea
- Diagnosis guide - 1 ea