Capnograph adopts mainstream method to measure the end-tidal carbon dioxide (EtCO2) and airway respiration rate (AwRR). It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow condition, as well as indirectly reflect the alveolar ventilation, which is applicable for use in emergency department, ICU, operating room, respiratory department and other occasions.
End-tidal CO2 measurement:
- Measurement range: 0 ~150mmHg
- Resolution: 1mmHg
Accuracy:
- 0~40 mmHg:±2 mmHg
- 41~70 mmHg:±5% of readings
- 71~100 mmHg:±8% of readings
- 101~150 mmHg:±10% of readings
Respiration rate measurement:
- Measurement range: 2rpm~150rpm
- Resolution: 1rpm
- Accuracy: ±1rpm
Safety characteristic:
- Device type: internally powered equipment
- Degree of protection against electric shock: type BF applied part
- Protection grade: IP22