SP950 is a syringe pump for veterinary use, and it adopts microcomputer system and microelectronic technology, features in high-intelligence and high-safety. LCD can display each parameter in real-time, with multi-function alarm, and alarm with visual, auditory and character prompt.
- Infusion rate and total volume can be set by user.
- Alarms for infusion pipe occlusion, syringe fall-off, communication abnormal, low battery.
- Bright, alarm volume and range of pressure alarm are adjustable.
- Enter KVO mode after infusion completed.
- The system will give prompt signal when there is no operation at power-on state.
- The device can equipped with disposable syringe (CNS), and accuracy can be ensured by adjusting parameters.
- Dimension: 310mm*140mm*135mm(L×W×H)
- Weight: 1.7kg
Performance
Specification requirement for syringe: 10ml/20ml/30ml/50ml (CNS)
Infusion rate:
- 10ml 0.1ml/h-100ml/h
- 20ml 0.1ml/h-200ml/h
- 30ml 0.1ml/h-300ml/h
- 50ml 0.1ml/h-500ml/h
Rate of quick feed:
- 10ml 100ml/h
- 20ml 200ml/h
- 30ml 300ml/h
- 50ml 500ml/h
Accumulated volume display: 0ml-999.9ml
Accuracy: ±3%
Pressure range for occlusion alarm:
Three grades are adjustable:
- Low: 40KPa ± 16KPa
- Middle: 70KPa ± 30KPa
- High: 100KPa± 35KPa
KVO rate: 0.1ml/h~5ml/h, adjustable
Safety classification: Class Ⅱ equipment, internally powered type BF
Maximal power consumption: 20VA~30VA, the device can continuously work for at least two hours with the injection rate of 5ml/h or 500ml/h after fully charged.
Power supply: AC100~240V 50/60Hz
Battery: rechargeable lithium battery, 7.4V, 3500mAh
Working environment:
Temperature: +5~40℃;
Atmospheric pressure: 860hPa~1060h;
Relative humidity: 20%~90%;
no strong vibration and no corrosive gas.
Package included
- 1x SP950 Syringe Pump
- 1x Power cord
- 1x Bracket
- 1x User manual